• A digital multimeter specialized for voltage measurement.
• DC voltage (V): 600 AC voltage (V): 600 DC current (A): - AC current (A): - Resistance (Ω): - Mass (g): 280
• Detection method: True RMS rectification (RMS) Maximum display: 6000 counts Width x Depth x Height: 82 x 50 x 161 mm Power source: AA batteries (R6P) x 2 (included) Resolution: DC voltage: 0.1 mV, AC voltage: 1 mV Range: Auto auto power off, data hold included
・Specialized in voltage measurement to prevent short circuit accidents
・Safe measurement with test cord with short-circuit prevention cap
・DC basic accuracy 0.5%
・True RMS value measurement that is resistant to distorted waveforms
・MIN/MAX function that can display minimum and maximum values
・Data hold, auto power off
<Specifications>
Measurement range/accuracy [DC voltage] = 600.0mV/6.000/60.00/600.0/600V (input impedance: 10MΩ, 600mV only 100MΩ), ±0.5%rdg±2dgt (600.0mV/6.000/60.00/600.0V),
±0.8%rdg±3dgt(600V)
[AC voltage]=6.000/60.00/600.0/600V (input impedance: 10MΩ), ±1.5%rdg±5dgt(6.000V), ±1.2%rdg±3dgt(60.00/600.0V), ±1.5%rdg±5dgt(600V)
Withstand voltage=AC3700V/1 minute
Compliant standard = IEC61010-1CATIII300V/CATII600V pollution degree 2,
IEC61010-031, IEC61326
Batteries used: AA batteries R6 (1.5V) x 2 (auto power off approximately 15 minutes)
External dimensions = 161(L)×82(W)×50(D)mm
Mass = approx. 280g (including battery)
Accessories = 7220A (measurement cord)